[Research on the Quality Control of Routine Reusable Pipeline and Disposable Pipeline of Ventilator].
Zhongguo Yi Liao Qi Xie Za Zhi
; 48(1): 111-113, 2024 Jan 30.
Article
en Zh
| MEDLINE
| ID: mdl-38384229
ABSTRACT
Objective:
To explore the effect of routine reusable pipeline and disposable pipeline on ventilator quality control results.Methods:
17 ventilators were randomly selected to conduct quality control using routine reusable pipeline and disposable pipeline respectively. Quality control data were recorded and then paired t-test method was used to analyze whether the difference between the two pipelines was significant or not.Results:
There were no significant differences in respiratory rate, tidal volume and end airway pressure between the two types of pipes ( Pï¼0.05). The airway peak pressure of routine reusable pipeline was significantly higher than disposable pipeline ( Pï¼0.05), but the difference was very small, only about 0.2 mbar which would not affect the conclusion of quality control.Conclusion:
Quality control of ventilator is not affected by routine reusable pipeline and disposable pipeline, which can be replaced by each other.Palabras clave