Effects of Nano-Electrospray Ionization Emitter Position on Unintentional In-Source Activation of Peptide and Protein Ions.
Shepherd, Samantha O; Green, Austin W; Resendiz, Elizabeth S; Newton, Kenneth R; Kurulugama, Ruwan T; Prell, James S.
Afiliación
  • Shepherd SO; Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States.
  • Green AW; Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States.
  • Resendiz ES; Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States.
  • Newton KR; Department of Chemistry and Biochemistry, University of Oregon, Eugene, Oregon 97403-1253, United States.
  • Kurulugama RT; Agilent Technologies, 5301 Stevens Creek Blvd, Santa Clara, California 95051, United States.
  • Prell JS; Agilent Technologies, 5301 Stevens Creek Blvd, Santa Clara, California 95051, United States.
J Am Soc Mass Spectrom ; 35(3): 498-507, 2024 Mar 06.
Article en En | MEDLINE | ID: mdl-38374644
ABSTRACT
Native ion mobility-mass spectrometry (IM-MS) typically introduces protein ions into the gas phase through nano-electrospray ionization (nESI). Many nESI setups have mobile stages for tuning the ion signal and extent of co-solute and salt adduction. However, tuning the position of the emitter capillary in nESI can have unintended downstream consequences for collision-induced unfolding or collision-induced dissociation (CIU/D) experiments. Here, we show that relatively small variations in the nESI emitter position can shift the midpoint (commonly called the "CID50" or "CIU50") potential of CID breakdown curves and CIU transitions by as much as 8 V on commercial instruments. A spatial "map" of the shift in CID50 for the loss of heme from holomyoglobin onto the emitter position on a Waters Synapt G2-Si mass spectrometer shows that emitter positions closer to the instrument inlet can result in significantly greater in-source activation, whereas different effects are found on an Agilent 6545XT instrument for the ions studied. A similar effect is observed for CID of the singly protonated leucine enkephalin peptide and Shiga toxin 1 subunit B homopentamer on the Waters Synapt G2-Si instrument. In-source activation effects on a Waters Synapt G2-Si are also investigated by examining the RMSD between CIU fingerprints acquired at different emitter positions and the shifts in CIU50 for structural transitions of bovine serum albumin and NIST monoclonal antibody.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Péptidos / Espectrometría de Masa por Ionización de Electrospray Idioma: En Revista: J Am Soc Mass Spectrom Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Péptidos / Espectrometría de Masa por Ionización de Electrospray Idioma: En Revista: J Am Soc Mass Spectrom Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos