GaN atomic electric fields from a segmented STEM detector: Experiment and simulation.
J Microsc
; 295(2): 140-146, 2024 Aug.
Article
en En
| MEDLINE
| ID: mdl-38372408
ABSTRACT
Atomic electric fields in a thin GaN sample are measured with the centre-of-mass approach in 4D-scanning transmission electron microscopy (4D-STEM) using a 12-segmented STEM detector in a Spectra 300 microscope. The electric fields, charge density and potential are compared to simulations and an experimental measurement using a pixelated 4D-STEM detector. The segmented detector benefits from a high recording speed, which enables measurements at low radiation doses. However, there is measurement uncertainty due to the limited number of segments analysed in this study.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
J Microsc
Año:
2024
Tipo del documento:
Article
País de afiliación:
Alemania
Pais de publicación:
Reino Unido