Evaluating a novel accelerated free-breathing late gadolinium enhancement imaging sequence for assessment of myocardial injury.
Magn Reson Imaging
; 108: 40-46, 2024 May.
Article
en En
| MEDLINE
| ID: mdl-38309379
ABSTRACT
INTRODUCTION:
Cardiac magnetic resonance imaging (MRI), including late gadolinium enhancement (LGE), plays an important role in the diagnosis and prognostication of ischemic and non-ischemic myocardial injury. Conventional LGE sequences require patients to perform multiple breath-holds and require long acquisition times. In this study, we compare image quality and assessment of myocardial LGE using an accelerated free-breathing sequence to the conventional standard-of-care sequence.METHODS:
In this prospective cohort study, a total of 41 patients post Coronavirus 2019 (COVID-19) infection were included. Studies were performed on a 1.5 Tesla scanner with LGE imaging acquired using a conventional inversion recovery rapid gradient echo (conventional LGE) sequence followed by the novel accelerated free-breathing (FB-LGE) sequence. Image quality was visually scored (ordinal scale from 1 to 5) and compared between conventional and free-breathing sequences using the Wilcoxon rank sum test. Presence of per-segment LGE was identified according to the American Heart Association 16-segment myocardial model and compared across both conventional LGE and FB-LGE sequences using a two-sided chi-square test. The perpatient LGE extent was also evaluated using both sequences and compared using the Wilcoxon rank sum test. Interobserver variability in detection of per-segment LGE and per-patient LGE extent was evaluated using Cohen's kappa statistic and interclass correlation (ICC), respectively.RESULTS:
The mean acquisition time for the FB-LGE sequence was 17 s compared to 413 s for the conventional LGE sequence (P < 0.001). Assessment of image quality was similar between both sequences (P = 0.19). There were no statistically significant differences in LGE assessed using the FB-LGE versus conventional LGE on a per-segment (P = 0.42) and per-patient (P = 0.06) basis. Interobserver variability in LGE assessment for FB-LGE was good for per-segment (= 0.71) and per-patient extent (ICC = 0.92) analyses.CONCLUSIONS:
The accelerated FB-LGE sequence performed comparably to the conventional standard-of-care LGE sequence in a cohort of patients post COVID-19 infection in a fraction of the time and without the need for breath-holding. Such a sequence could impact clinical practice by increasing cardiac MRI throughput and accessibility for frail or acutely ill patients unable to perform breath-holding.Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Medios de Contraste
/
COVID-19
Tipo de estudio:
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Humans
Idioma:
En
Revista:
Magn Reson Imaging
Año:
2024
Tipo del documento:
Article
País de afiliación:
Canadá
Pais de publicación:
Países Bajos