LETd Optimization Verification With an SOI Microdosimeter.
Int J Radiat Oncol Biol Phys
; 119(3): 946-956, 2024 Jul 01.
Article
en En
| MEDLINE
| ID: mdl-38300188
ABSTRACT
PURPOSE:
A first of its kind experimental verification of dose-averaged linear energy transfer (LETd) optimized treatment plans for proton therapy has been carried out using a silicon-on-insulator microdosimeter at the Massachusetts General Hospital (MGH), Boston, USA. METHODS AND MATERIALS Three clinical treatment plans of a typical ependymoma structure set were designed using the standard clinical approach, the proposed protocol approach, and a one-field approach. The plans were then reoptimized to reduce the LETd-weighted dose in the brain stem. All six plans were delivered in a solid water phantom and the experimental yDâ¾ measured.RESULTS:
After LETd optimization, a reduction in yDâ¾ was found within the brain stem by an average of 12%, 19%, and 4% for the clinical, protocol, and one-field plans, respectively, while maintaining adequate coverage of the tumor structure. The experimental LETd-weighted doses were in agreement with the treatment planning system calculations and Monte Carlo simulations and reinforced the improvement of the optimization.CONCLUSIONS:
This work demonstrates the first experimental verification of the clinical implementation of LETd optimization for patient treatment with proton therapy.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Dosificación Radioterapéutica
/
Planificación de la Radioterapia Asistida por Computador
/
Neoplasias Encefálicas
/
Método de Montecarlo
/
Transferencia Lineal de Energía
/
Fantasmas de Imagen
/
Ependimoma
/
Terapia de Protones
Tipo de estudio:
Guideline
Límite:
Humans
Idioma:
En
Revista:
Int J Radiat Oncol Biol Phys
Año:
2024
Tipo del documento:
Article
Pais de publicación:
Estados Unidos