Noninvasive monitoring of the vagus nerve during thyroid surgery using cutaneous adhesive and needle electrodes: What is the optimal configuration?
Karcioglu, Amanda Silver; Trinh, Lily N; Mcllroy, Dioan; Okose, Okenwa C; Wang, Bo; Behr, Ian J; Cheung, Anthony Y; Srikanthan, Adithya; Russell, Marika D; Kamani, Dipti; Abdelhamid Ahmed, Amr H; Randolph, Gregory W.
Afiliación
  • Karcioglu AS; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Trinh LN; Division of Otolaryngology - Head and Neck Surgery, Department of Surgery, NorthShore University Health System, Evanston, Illinois, USA.
  • Mcllroy D; Pritzker School of Medicine, The University of Chicago, Chicago, Illinois, USA.
  • Okose OC; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Wang B; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Behr IJ; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Cheung AY; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Srikanthan A; Department of Thyroid Surgery, Fujian Medical University Union Hospital, Fuzhou, China.
  • Russell MD; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Kamani D; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Abdelhamid Ahmed AH; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
  • Randolph GW; Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology - Head and Neck Surgery, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
Head Neck ; 46(8): 1893-1901, 2024 08.
Article en En | MEDLINE | ID: mdl-38294128
ABSTRACT

OBJECTIVE:

Endotracheal tube (ETT) surface electrodes are used to monitor the vagus nerve (VN), recurrent laryngeal nerve (RLN), and external branch of the superior laryngeal nerve (EBSLN) during thyroid and parathyroid surgery. Alternative nerve monitoring methods are desirable when intubation under general anesthesia is not desirable or possible. In this pilot study, we compared the performance of standard ETT electrodes to four different noninvasive cutaneous recording electrode types (two adhesive electrodes and two needle electrodes) in three different orientations.

METHODS:

The VN was stimulated directly during thyroid and parathyroid surgery using a Prass stimulator probe. Electromyographic (EMG) responses for each patient were recorded using an ETT plus one of the following four cutaneous electrode types large-foot adhesive, small-foot adhesive, long-needle and short-needle. Each of the four electrode types was placed in three orientations (1) bilateral, (2) ipsilateral mediolateral, and (3) ipsilateral craniocaudal.

RESULTS:

Four surgical cases were utilized for data collection with the repetitive measures obtained in each subject. Bilateral electrode orientation was superior to ipsilateral craniocaudal and ipsilateral mediolateral orientations. Regardless of electrodes type, all amplitudes in the bilateral orientation were >100 µV. When placed bilaterally, the small-foot adhesive and the long-needle electrodes obtained the highest EMG amplitudes as a percentage of ETT amplitudes.

CONCLUSION:

Cutaneous electrodes could potentially be used to monitor the VN during thyroid and parathyroid procedures. Different electrode types vary in their ability to record amplitudes and latencies. Bilateral orientation improves EMG responses in all electrode types. Additional validation of cutaneous electrodes as an alternative noninvasive method to monitor the VN is needed.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tiroidectomía / Nervio Vago / Electrodos / Electromiografía / Agujas Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Head Neck Asunto de la revista: NEOPLASIAS Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tiroidectomía / Nervio Vago / Electrodos / Electromiografía / Agujas Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Head Neck Asunto de la revista: NEOPLASIAS Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos