The investigation of dose rate and photon beam energy dependence of optimized PASSAG polymer gel dosimeter using magnetic resonance imaging.
J Xray Sci Technol
; 32(3): 751-764, 2024.
Article
en En
| MEDLINE
| ID: mdl-38217634
ABSTRACT
OBJECTIVE:
It seems that dose rate (DR) and photon beam energy (PBE) may influence the sensitivity and response of polymer gel dosimeters. In the current project, the sensitivity and response dependence of optimized PASSAG gel dosimeter (OPGD) on DR and PBE were assessed. MATERIALS ANDMETHODS:
We fabricated the OPGD and the gel samples were irradiated with various DRs and PBEs. Then, the sensitivity and response (R2) of OPGD were obtained by MRI at various doses and post-irradiation times.RESULTS:
Our analysis showed that the sensitivity and response of OPGD are not affected by the evaluated DRs and PBEs. It was also found that the dose resolution values of OPGD ranged from 9 to 33 cGy and 12 to 34 cGy for the evaluated DRs and PBEs, respectively. Additionally, the data demonstrated that the sensitivity and response dependence of OPGD on DR and PBE do not vary over various times after the irradiation.CONCLUSIONS:
The findings of this research project revealed that the sensitivity and response dependence of OPGD are independent of DR and PBE.Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Imagen por Resonancia Magnética
/
Fotones
/
Dosímetros de Radiación
/
Geles
Idioma:
En
Revista:
J Xray Sci Technol
Asunto de la revista:
RADIOLOGIA
Año:
2024
Tipo del documento:
Article
País de afiliación:
China
Pais de publicación:
Países Bajos