Relationship Between Achilles Tendon Stiffness Using Myoton PRO and Translation Using a Tensile Testing Machine: A Biomechanical Study of a Porcine Model.
Kurashina, Wataru; Takahashi, Tsuneari; Sasanuma, Hideyuki; Saitsu, Akihiro; Takeshita, Katsushi.
Afiliación
  • Kurashina W; Graduate School of Medicine, Jichi Medical University, Shimotsuke, JPN.
  • Takahashi T; Department of Rehabilitation, Tochigi Medical Center Shimotsuga, Tochigi, JPN.
  • Sasanuma H; Department of Orthopaedic Surgery, Ishibashi General Hospital, Shimotsuke, JPN.
  • Saitsu A; Department of Orthopaedics, Jichi Medical University, Shimotsuke, JPN.
  • Takeshita K; Department of Orthopaedics, Jichi Medical University, Shimotsuke, JPN.
Cureus ; 15(11): e49359, 2023 Nov.
Article en En | MEDLINE | ID: mdl-38146575
ABSTRACT
Background Achilles tendinopathy is a common ankle disorder in both the general population and athletes. This condition can alter the mechanical characteristics of the Achilles tendon (AT) by decreasing tendon stiffness. Achilles tendinopathy is primarily treated conservatively; however, few monitoring tools exist for evaluating the condition of the AT. The Myoton PRO (Myoton AS, Tallinn, Estonia) device is a handheld tool used to evaluate tissue stiffness. However, no basic studies have examined the validity of Myoton PRO for assessing the AT. This study aimed to assess the validity of Myoton PRO using animal ATs and to examine its clinical applicability. Methods We used 28 fresh porcine ankles and evaluated AT stiffness at the calcaneus insertion site (AT0) and 2.0 cm above the calcaneus (AT2) using Myoton PRO. We also measured changes in the AT length using a tensile testing machine during the cyclic loading test. We investigated the correlation between dynamic stiffness and length change. Furthermore, we assessed the difference in stiffness between AT0 and AT2. Results The dynamic stiffness was 717.6 ± 183.1 N/m at AT0 and 467.4 ± 152.3 N/m at AT2. The change in length during the cyclic loading test was 1.8 ± 0.7 mm. The correlation between dynamic stiffness and length change was as follows AT0, r=-0.61; AT2, r=-0.64 (P<0.001). The dynamic stiffness at AT0 was significantly greater than that at AT2 (P<0.001). Conclusions AT assessment using Myoton PRO has potential clinical utility as an indicator of tissue stiffness.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Cureus Año: 2023 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Cureus Año: 2023 Tipo del documento: Article Pais de publicación: Estados Unidos