Comment on "So you think you can DAS? A viewpoint on delay-and-sum beamforming" [Ultrasonics 111 (2021) 106309].
Schiffner, Martin F; Schmitz, Georg.
Afiliación
  • Schiffner MF; Chair of Medical Engineering, Ruhr University Bochum, 44780 Bochum, Germany. Electronic address: martin.schiffner@rub.de.
  • Schmitz G; Chair of Medical Engineering, Ruhr University Bochum, 44780 Bochum, Germany. Electronic address: georg.schmitz@rub.de.
Ultrasonics ; 138: 107221, 2024 Mar.
Article en En | MEDLINE | ID: mdl-38134514
ABSTRACT
It is shown herein that Perrot et al., who reviewed delay-and-sum beamforming for ultrafast ultrasound imaging in [Ultrasonics 111 (2021) 106309], misinterpreted the purpose of dynamic receive apertures. Such apertures widen with the focal length as a function of a given f-number and improve the image quality by suppressing grating lobes. Perrot et al., however, attributed erroneously the image quality improvement to suppression of measurement noise and, in doing so, proposed a suboptimal method to determine an f-number.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Ultrasonics Año: 2024 Tipo del documento: Article Pais de publicación: Países Bajos