Video analysis of optic-haptic-interaction during hydrophobic acrylic intraocular lens implantation using preloaded injectors.
Yan, Weijia; Borkenstein, Andreas F; Khoramnia, Ramin; Borkenstein, Eva-Maria; Auffarth, Gerd U.
Afiliación
  • Yan W; The David J Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
  • Borkenstein AF; Borkenstein and Borkenstein, Private Practice at Privatklinik Kreuzschwestern, Graz, Austria.
  • Khoramnia R; The David J Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
  • Borkenstein EM; Borkenstein and Borkenstein, Private Practice at Privatklinik Kreuzschwestern, Graz, Austria.
  • Auffarth GU; The David J Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. Gerd.Auffarth@med.uni-heidelberg.de.
BMC Ophthalmol ; 23(1): 515, 2023 Dec 19.
Article en En | MEDLINE | ID: mdl-38115049
ABSTRACT

OBJECTIVE:

To compare the optic-haptic interaction of different hydrophobic acrylic IOLs after using six preloaded injectors.

METHODS:

We reviewed the video-recordings of procedures on a total of 388 eyes that underwent phacoemulsification and intraocular lens (IOL) implantation. For six preloaded injectors multiSert (Hoya Surgical Optics) [System 1], TECNIS Simplicity (Johnson & Johnson Vision) [System 2], TECNIS iTec (Johnson & Johnson Vision) [System 3], AutonoMe (Alcon, Laboratories) [System 4], Bluesert (Carl Zeiss Meditec) [System 5], and Prosert (OphthalmoPro GmbH) [System 6], we noted in each case the time of IOL delivery and made a descriptive observation of IOL insertion and optic-haptic-interaction.

RESULTS:

We defined standard haptic behavior where the haptics emerged "folded" from the injector and quickly recovered their pre-implantation appearance. The incidence where the leading haptic emerged in a deformed way for System 1 was 20%, System 2 19%, System 3 14%, System 4 56%, System 5 24% and System 6 5%. For trailing haptic deformed behavior, the incidence was 36%, 6%, 4%, 8%, 18% and 2%, respectively for Systems 1 to 6. Optic-haptic adhesion occurred in 2% of cases for System 1, 44% for System 2, 52% for System 3, 48% for System 4, and 11% for System 6 (P < 0.05). Adhesion was not found with System 5.

CONCLUSIONS:

We observed different deformed behavior for leading and trailing haptics in the six preloaded systems, some systems had as much as 52% optic-haptic adhesion.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Extracción de Catarata / Facoemulsificación / Cápsula del Cristalino / Lentes Intraoculares Límite: Humans Idioma: En Revista: BMC Ophthalmol Asunto de la revista: OFTALMOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Alemania Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Extracción de Catarata / Facoemulsificación / Cápsula del Cristalino / Lentes Intraoculares Límite: Humans Idioma: En Revista: BMC Ophthalmol Asunto de la revista: OFTALMOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Alemania Pais de publicación: Reino Unido