Superior bone fixation of conical compared with hemispherical trapezial cup design: an experimental radiostereometry study.
Stilling, Maiken; Dremstrup, Lene; Hansen, Torben Bæk; Thillemann, Janni Kjærgaard.
Afiliación
  • Stilling M; Department of Orthopaedic Surgery, Aarhus University Hospital, Palle Juul-Jensens Boulevard 165 J801, DK- 8200, Aarhus N, Denmark. maiken.stilling@clin.au.dk.
  • Dremstrup L; Department of Clinical Medicine, Aarhus University, Aarhus N, Denmark. maiken.stilling@clin.au.dk.
  • Hansen TB; AutoRSA Research Group, Orthopaedic Research Unit, Aarhus University Hospital, Aarhus N, Denmark. maiken.stilling@clin.au.dk.
  • Thillemann JK; Department of Orthopaedics, University Clinic for Hand, Hip and Knee Surgery, Gødstrup Hospital, Herning, Denmark. maiken.stilling@clin.au.dk.
J Exp Orthop ; 10(1): 127, 2023 Nov 30.
Article en En | MEDLINE | ID: mdl-38032446
PURPOSE: The most used cup designs for trapeziometacarpal (TMC) arthroplasty are of hemispherical and conical geometrical shape. Using a validated pig bone model, we compared the bone fixation using radiostereometry (RSA). METHODS: Twenty saddle-shaped pig forefoot bones were prepared with insertion of bone markers and reaming. Hemispherical Type T cups (Beznoska, Kladno, Czech Republic) (N = 10) and conical Moovis cups (Stryker, Pusignan, France) (N = 10) of 9-mm diameter were inserted press-fit. The bones were fixed in cement blocks for stability, and the cups were loaded in a motorized test stand. First, a low-pressure cyclic load test (0-150N) with 130 compression cycles was performed. Next, a push-in test of progressive loads with 50N increments (range: 150-900N) was applied until a visual change in cup position appeared. Cup migration was evaluated with RSA after every new load application. Cup failure was defined as total translation > 0.5 mm between two load applications. RESULTS: Both cup types tolerated a compression load of 450 N without failure. Beyond this load level, the total translation cup migration of mean 0.20 mm (95% CI 0.11; 0.30) for the Type T group was higher than mean 0.10 mm (95% CI 0.06; 0.15) of the Moovis group (p = 0.046). The Hazard ratio for failure was 0.52 (95% CI 0.12; 2.17) (p = 0.37), indicating that the risk of failure was two-fold higher in the Type T group. CONCLUSION: We conclude that conical TMC cups have superior fixation as compared to hemispherical cups above a loading level of 450 N, which correspond to a 3.8 kg tip-pinch. In a clinical perspective, based on the fixation strength of both cup types, it seems safe to allow light-load activities of daily living such as buttoning a shirt and using a key shortly after surgery and until sufficient osseointegration is achieved.