Endovascular retrieval of a fractured Optease inferior vena cava filter using endobronchial forceps and intraoperative cone-beam computed tomography guidance.
J Vasc Surg Cases Innov Tech
; 9(3): 101187, 2023 Sep.
Article
en En
| MEDLINE
| ID: mdl-37799830
Endovascular retrieval of fractured inferior vena cava (IVC) filters after the manufacturer recommended indwelling time can be challenging and require advanced retrieval techniques. We describe an endovascular retrieval technique of a fractured Optease IVC filter in a 57-year-old woman using endobronchial forceps and intraoperative cone-beam computed tomography guidance. Following incomplete filter retrieval, the location and orientation of fractured strut was confirmed by cone-beam computed tomography venography. The embedded filter fragment was then successfully removed using endobronchial forceps via a transjugular venous approach. In the present report, we highlight the additional value of intraoperative cross-sectional imaging, in conjunction with advanced endovascular techniques, for retrieval of challenging IVC filters.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Tipo de estudio:
Guideline
Idioma:
En
Revista:
J Vasc Surg Cases Innov Tech
Año:
2023
Tipo del documento:
Article
Pais de publicación:
Estados Unidos