Lack of Consensus Between Measurements of Plasma Phylloquinone by Enzyme-Linked Immunosorbent assays and a Well-Validated High-Performance Liquid Chromatographic Method.
Curr Dev Nutr
; 7(7): 101959, 2023 Jul.
Article
en En
| MEDLINE
| ID: mdl-37408980
Enzyme-linked i2mmunosorbent assays (ELISAs) that measure circulating phylloquinone have become commercially available, but their validity is uncertain. The objective of this study was to compare plasma phylloquinone concentrations measured using two commercially available ELISAs with concentrations measured using a validated high-performance liquid chromatography (HPLC) assay in 108 samples obtained from participants in a depletion (â¼10 mcg phylloquinone/d)-supplementation (â¼500 mcg phylloquinone/d) study. The geometric mean of plasma phylloquinone measured with ELISA A was 0.70 nmol/L, 37% lower than that measured with HPLC. The mean of the ELISA B measures was 12.4 nmol/L, >700% higher than the HPLC measures. Plasma phylloquinone measured using HPLC was significantly lower during phylloquinone depletion than supplementation (0.4 ± 0.1 compared with 1.2 ± 0.2 nmol/L; P < 0.001). Neither of the two ELISAs detected any significant difference in plasma phylloquinone concentrations between depletion and supplementation (ELISA A, P = 0.76; ELISA B, P = 0.29). These findings reinforce the need to validate plasma phylloquinone assays as they become available. Curr Dev Nutr 2023;x:xx.