Fracture of Contemporary Femoral Stems: Common Trends in This Rare Occurrence.
Cook, Stephen D; Patron, Laura P; Lavernia, Carlos J; Gibian, Joseph T; Hong, Thomas S; Bendich, Ilya.
Afiliación
  • Cook SD; Fellowship of Orthopaedic Researchers, Metairie, Louisiana.
  • Patron LP; Fellowship of Orthopaedic Researchers, Metairie, Louisiana.
  • Lavernia CJ; Lavernia Orthopaedic Surgery, Miami, Florida.
  • Gibian JT; Washington University in St. Louis School of Medicine, St. Louis, Missouri.
  • Hong TS; Washington University in St. Louis School of Medicine, St. Louis, Missouri.
  • Bendich I; Washington University in St. Louis School of Medicine, St. Louis, Missouri.
J Arthroplasty ; 38(7S): S285-S291, 2023 07.
Article en En | MEDLINE | ID: mdl-37086930
BACKGROUND: Fracture of contemporary femoral stems is a rare occurrence in total hip arthroplasty. A knowledge gap remains regarding manufacturing, patient, and surgeon factors that may contribute to the increased risk of this complication. METHODS: We analyzed 13 contemporary fractured porous-coated femoral stems of various designs to determine cause and contributing factors of mechanical failure. Cases included 12 men and 1 woman who had an average age at index surgery of 53 years (range, 34 to 76 years). There were 10 of 13 patients who had a body mass index more than 30 (obese); 3 of the 10 had a body mass index more than 40. The mean time to fracture was 7.6 years (range, 7 months to 12 years). RESULTS: There were 4 titanium alloy stems that fractured an average of 3.6 years postrevision surgery for head/cup exchange and had associated iatrogenic mechanical and electrocautery damage to the femoral neck at fracture initiation sites. There were 6 modular stems that failed at the stem-sleeve or stem-neck interfaces with evidence of fretting corrosion. For 2 stem-neck fractures, mismatched head/stem combinations from different manufacturers resulted in untested mechanical offsets and loading. There were 2 proximal neck fractures and 1 mid-shaft fracture of coated cobalt-chromium alloy stems that occurred in 3 obese men. The neck fractures (10 to 12 years) were well-fixed stems. Lack of proximal fixation contributed to the mid-shaft fracture (7 months). CONCLUSION: While rare, femoral stem fractures pose catastrophic outcomes in primary and revision total hip arthroplasty. Manufacturing, patient, and surgical factors contributing to stem failures were identified, including patient obesity, heat-treatment reduction of mechanical properties, iatrogenic implant damage, and mixing of different vendor stems and heads.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Fracturas del Fémur / Prótesis de Cadera Tipo de estudio: Etiology_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2023 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Fracturas del Fémur / Prótesis de Cadera Tipo de estudio: Etiology_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2023 Tipo del documento: Article Pais de publicación: Estados Unidos