Performance of Contactless Respiratory Rate Monitoring by Albus HomeTM, an Automated System for Nocturnal Monitoring at Home: A Validation Study.
Do, William; Russell, Richard; Wheeler, Christopher; Lockwood, Megan; De Vos, Maarten; Pavord, Ian; Bafadhel, Mona.
Afiliación
  • Do W; Albus Health, Oxford OX4 2HN, UK.
  • Russell R; Respiratory Medicine Unit, Nuffield Department of Medicine, University of Oxford, Oxford OX3 7FZ, UK.
  • Wheeler C; Royal Brompton Hospital, London SW3 6NP, UK.
  • Lockwood M; Albus Health, Oxford OX4 2HN, UK.
  • De Vos M; Department of Electrical Engineering and Department of Development and Regeneration, KU Leuven, 3000 Leuven, Belgium.
  • Pavord I; Respiratory Medicine Unit, Nuffield Department of Medicine, University of Oxford, Oxford OX3 7FZ, UK.
  • Bafadhel M; King's Centre for Lung Health, School of Immunology and Microbial Sciences, Faculty of Life Sciences and Medicine, King's College London, London SE1 1UL, UK.
Sensors (Basel) ; 22(19)2022 Sep 21.
Article en En | MEDLINE | ID: mdl-36236241
Respiratory rate (RR) is a clinically important predictor of cardio-respiratory deteriorations. The mainstay of clinical measurement comprises the manual counting of chest movements, which is variable between clinicians and limited to sporadic readings. Emerging solutions are limited by poor adherence and acceptability or are not clinically validated. Albus HomeTM is a contactless and automated bedside system for nocturnal respiratory monitoring that overcomes these limitations. This study aimed to validate the accuracy of Albus Home compared to gold standards in real-world sleeping environments. Participants undertook overnight monitoring simultaneously using Albus Home and gold-standard polygraphy with thoraco-abdominal respiratory effort belts (SomnomedicsEU). Reference RR readings were obtained by clinician-count of polygraphy data. For both the Albus system and reference, RRs were measured in 30-s segments, reported as breaths/minute, and compared. Accuracy was defined as the percentage of RRs from the Albus system within ±2 breaths/minute of reference counts. Across a diverse validation set of 32 participants, the mean accuracy exceeded 98% and was maintained across different participant characteristics. In a Bland-Altman analysis, Albus RRs had strong agreement with reference mean differences and the limits of agreement of -0.4 and ±1.2 breaths/minute, respectively. Albus Home is a contactless yet accurate system for automated respiratory monitoring. Validated against gold -standard methods, it enables long-term, reliable nocturnal monitoring without patient burden.