Exploiting a strong coupling regime of organic pentamer surface plasmon resonance based on the Otto configuration for creatinine detection.
Opt Express
; 30(9): 14478-14491, 2022 Apr 25.
Article
en En
| MEDLINE
| ID: mdl-35473189
The sandwiched material-analyte layer in the surface plasmon resonance (SPR)-Otto configuration emulates an optical cavity and, coupled with large optical nonlinearity material, the rate of light escaping from the system is reduced, allowing the formation of a strong coupling regime. Here, we report an organic pentamer SPR sensor using the Otto configuration to induce a strong coupling regime for creatinine detection. Prior to that, the SPR sensor chip was modified with an organic pentamer, 1,4-bis[2-(5-thiophene-2-yl)-1-benzothiopene]-2,5-dioctyloxybenzene (BOBzBT2). To improve the experimental calibration curve, a normalisation approach based on the strong coupling-induced second dip was also developed. By using this procedure, the performance of the sensor improved to 0.11 mg/dL and 0.36 mg/dL for the detection and quantification limits, respectively.