Robot-Assisted Stereotaxy Reduces Target Error: A Meta-Analysis and Meta-Regression of 6056 Trajectories.
Philipp, Lucas R; Matias, Caio M; Thalheimer, Sara; Mehta, Shyle H; Sharan, Ashwini; Wu, Chengyuan.
Afiliación
  • Philipp LR; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
  • Matias CM; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
  • Thalheimer S; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
  • Mehta SH; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
  • Sharan A; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
  • Wu C; Department of Neurological Surgery, Thomas Jefferson University Hospitals, Philadelphia, Pennsylvania.
Neurosurgery ; 88(2): 222-233, 2021 01 13.
Article en En | MEDLINE | ID: mdl-33045739
BACKGROUND: The pursuit of improved accuracy for localization and electrode implantation in deep brain stimulation (DBS) and stereoelectroencephalography (sEEG) has fostered an abundance of disparate surgical/stereotactic practices. Specific practices/technologies directly modify implantation accuracy; however, no study has described their respective influence in multivariable context. OBJECTIVE: To synthesize the known literature to statistically quantify factors affecting implantation accuracy. METHODS: A systematic review and meta-analysis was conducted to determine the inverse-variance weighted pooled mean target error (MTE) of implanted electrodes among patients undergoing DBS or sEEG. MTE was defined as Euclidean distance between planned and final electrode tip. Meta-regression identified moderators of MTE in a multivariable-adjusted model. RESULTS: A total of 37 eligible studies were identified from a search return of 2,901 potential articles (2002-2018) - 27 DBS and 10 sEEG. Random-effects pooled MTE = 1.91 mm (95% CI: 1.7-2.1) for DBS and 2.34 mm (95% CI: 2.1-2.6) for sEEG. Meta-regression identified study year, robot use, frame/frameless technique, and intraoperative electrophysiologic testing (iEPT) as significant multivariable-adjusted moderators of MTE (P < .0001, R2 = 0.63). Study year was associated with a 0.92-mm MTE reduction over the 16-yr study period (P = .0035), and robot use with a 0.79-mm decrease (P = .0019). Frameless technique was associated with a mean 0.50-mm (95% CI: 0.17-0.84) increase, and iEPT use with a 0.45-mm (95% CI: 0.10-0.80) increase in MTE. Registration method, imaging type, intraoperative imaging, target, and demographics were not significantly associated with MTE on multivariable analysis. CONCLUSION: Robot assistance for stereotactic electrode implantation is independently associated with improved accuracy and reduced target error. This remains true regardless of other procedural factors, including frame-based vs frameless technique.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Robótica / Técnicas Estereotáxicas / Imagenología Tridimensional / Estimulación Encefálica Profunda / Electrocorticografía Tipo de estudio: Prognostic_studies / Systematic_reviews Límite: Female / Humans Idioma: En Revista: Neurosurgery Año: 2021 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Robótica / Técnicas Estereotáxicas / Imagenología Tridimensional / Estimulación Encefálica Profunda / Electrocorticografía Tipo de estudio: Prognostic_studies / Systematic_reviews Límite: Female / Humans Idioma: En Revista: Neurosurgery Año: 2021 Tipo del documento: Article Pais de publicación: Estados Unidos