Feasibility of a GATE Monte Carlo platform in a clinical pretreatment QA system for VMAT treatment plans using TrueBeam with an HD120 multileaf collimator.
J Appl Clin Med Phys
; 20(10): 101-110, 2019 Oct.
Article
en En
| MEDLINE
| ID: mdl-31544350
PURPOSE: To evaluate the quality of patient-specific complicated treatment plans, including commercialized treatment planning systems (TPS) and commissioned beam data, we developed a process of quality assurance (QA) using a Monte Carlo (MC) platform. Specifically, we constructed an interface system that automatically converts treatment plan and dose matrix data in digital imaging and communications in medicine to an MC dose-calculation engine. The clinical feasibility of the system was evaluated. MATERIALS AND METHODS: A dose-calculation engine based on GATE v8.1 was embedded in our QA system and in a parallel computing system to significantly reduce the computation time. The QA system automatically converts parameters in volumetric-modulated arc therapy (VMAT) plans to files for dose calculation using GATE. The system then calculates dose maps. Energies of 6 MV, 10 MV, 6 MV flattening filter free (FFF), and 10 MV FFF from a TrueBeam with HD120 were modeled and commissioned. To evaluate the beam models, percentage depth dose (PDD) values, MC calculation profiles, and measured beam data were compared at various depths (Dmax , 5 cm, 10 cm, and 20 cm), field sizes, and energies. To evaluate the feasibility of the QA system for clinical use, doses measured for clinical VMAT plans using films were compared to dose maps calculated using our MC-based QA system. RESULTS: A LINAC QA system was analyzed by PDD and profile according to the secondary collimator and multileaf collimator (MLC). Values for MC calculations and TPS beam data obtained using CC13 ion chamber (IBA Dosimetry, Germany) were consistent within 1.0%. Clinical validation using a gamma index was performed for VMAT treatment plans using a solid water phantom and arbitrary patient data. The gamma evaluation results (with criteria of 3%/3 mm) were 98.1%, 99.1%, 99.2%, and 97.1% for energies of 6 MV, 10 MV, 6 MV FFF, and 10 MV FFF, respectively. CONCLUSIONS: We constructed an MC-based QA system for evaluating patient treatment plans and evaluated its feasibility in clinical practice. We observed robust agreement between dose calculations from our QA system and measurements for VMAT plans. Our QA system could be useful in other clinical settings, such as small-field SRS procedures or analyses of secondary cancer risk, for which dose calculations using TPS are difficult to verify.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Aceleradores de Partículas
/
Garantía de la Calidad de Atención de Salud
/
Planificación de la Radioterapia Asistida por Computador
/
Método de Montecarlo
/
Fantasmas de Imagen
/
Radioterapia de Intensidad Modulada
Tipo de estudio:
Health_economic_evaluation
/
Prognostic_studies
Límite:
Humans
Idioma:
En
Revista:
J Appl Clin Med Phys
Asunto de la revista:
BIOFISICA
Año:
2019
Tipo del documento:
Article
Pais de publicación:
Estados Unidos