A DuMond-type crystal spectrometer for synchrotron-based X-ray emission studies in the energy range of 15-26 keV.
Jagodzinski, P; Szlachetko, J; Dousse, J-Cl; Hoszowska, J; Szlachetko, M; Vogelsang, U; Banas, D; Pakendorf, T; Meents, A; van Bokhoven, J A; Kubala-Kukus, A; Pajek, M; Nachtegaal, M.
Afiliación
  • Jagodzinski P; Swiss Light Source, Paul Scherrer Institute (PSI), CH-5232 Villigen, Switzerland.
  • Szlachetko J; Institute of Nuclear Physics Polish Academy of Science, PL-31-342 Kraków, Poland.
  • Dousse JC; Department of Physics, University of Fribourg, CH-1700 Fribourg, Switzerland.
  • Hoszowska J; Department of Physics, University of Fribourg, CH-1700 Fribourg, Switzerland.
  • Szlachetko M; Swiss Light Source, Paul Scherrer Institute (PSI), CH-5232 Villigen, Switzerland.
  • Vogelsang U; Swiss Light Source, Paul Scherrer Institute (PSI), CH-5232 Villigen, Switzerland.
  • Banas D; Institute of Physics, Jan Kochanowski University, PL-25-406 Kielce, Poland.
  • Pakendorf T; PETRA, Deutsches Elektronen-Synchrotron (DESY), DE-22607 Hamburg, Germany.
  • Meents A; PETRA, Deutsches Elektronen-Synchrotron (DESY), DE-22607 Hamburg, Germany.
  • van Bokhoven JA; Swiss Light Source, Paul Scherrer Institute (PSI), CH-5232 Villigen, Switzerland.
  • Kubala-Kukus A; Institute of Physics, Jan Kochanowski University, PL-25-406 Kielce, Poland.
  • Pajek M; Institute of Physics, Jan Kochanowski University, PL-25-406 Kielce, Poland.
  • Nachtegaal M; Swiss Light Source, Paul Scherrer Institute (PSI), CH-5232 Villigen, Switzerland.
Rev Sci Instrum ; 90(6): 063106, 2019 Jun.
Article en En | MEDLINE | ID: mdl-31255015
The design and performance of a high-resolution transmission-type X-ray spectrometer for use in the 15-26 keV energy range at synchrotron light sources is reported. Monte Carlo X-ray-tracing simulations were performed to optimize the performance of the transmission-type spectrometer, based on the DuMond geometry, for use at the Super X-ray absorption beamline of the Swiss Light Source at the Paul Scherrer Institute. This spectrometer provides an instrumental energy resolution of 3.5 eV for X-ray emission lines around 16 keV and 12.5 eV for emission lines at 26 keV, which is comparable to the natural linewidths of the K and L X-ray transitions in the covered energy range. First experimental data are presented and compared with results of the Monte Carlo X-ray simulations.