Water-equivalence of gel dosimeters for radiology medical imaging.
Valente, M; Vedelago, J; Chacón, D; Mattea, F; Velásquez, J; Pérez, P.
Afiliación
  • Valente M; Instituto de Física Enrique Gaviola - CONICET, Av. M. Allende s/n, 5000 Córdoba, Argentina; Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes por Rayos X, Universidad Nacional de Córdoba, Av. M. Allende s/n, 5000 Córdoba, Argentina; Centro de Física e Ingenie
  • Vedelago J; Instituto de Física Enrique Gaviola - CONICET, Av. M. Allende s/n, 5000 Córdoba, Argentina; Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes por Rayos X, Universidad Nacional de Córdoba, Av. M. Allende s/n, 5000 Córdoba, Argentina.
  • Chacón D; Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes por Rayos X, Universidad Nacional de Córdoba, Av. M. Allende s/n, 5000 Córdoba, Argentina; Departamento de Física, Universidad Nacional, Heredia, Costa Rica, Avenida 1, Calle 9. Apartado Postal: 86-3000, Hered
  • Mattea F; Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes por Rayos X, Universidad Nacional de Córdoba, Av. M. Allende s/n, 5000 Córdoba, Argentina; Universidad Nacional de Córdoba. Facultad de Ciencias Químicas. Departamento de Química Orgánica, Córdoba, Argentina;
  • Velásquez J; ICOS Inmunomédica, Lago Puyehue, 1750 Temuco, Chile; Facultad de Ciencias, Escuela de Tecnología Médica, Universidad Mayor, Chile.
  • Pérez P; Instituto de Física Enrique Gaviola - CONICET, Av. M. Allende s/n, 5000 Córdoba, Argentina; Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes por Rayos X, Universidad Nacional de Córdoba, Av. M. Allende s/n, 5000 Córdoba, Argentina.
Appl Radiat Isot ; 141: 193-198, 2018 Nov.
Article en En | MEDLINE | ID: mdl-29555376
International dosimetry protocols are based on determinations of absorbed dose to water. Ideally, the phantom material should be water equivalent; that is, it should have the same absorption and scatter properties as water. This study presents theoretical, experimental and Monte Carlo modeling of water-equivalence of Fricke and polymer (NIPAM, PAGAT and itaconic acid ITABIS) gel dosimeters. Mass and electronic densities along with effective atomic number were calculated by means of theoretical approaches. Samples were scanned by standard computed tomography. Photon mass attenuation coefficients and electron stopping powers were examined. Theoretical, Monte Carlo and experimental results confirmed good water-equivalence for all gel dosimeters. Overall variations with respect to water in the low energy radiology range (up to 130 kVp) were found to be less than 3% in average.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosímetros de Radiación Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2018 Tipo del documento: Article Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosímetros de Radiación Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2018 Tipo del documento: Article Pais de publicación: Reino Unido