Radiation dose optimisation for conventional imaging in infants and newborns using automatic dose management software: an application of the new 2013/59 EURATOM directive.
Alejo, L; Corredoira, E; Sánchez-Muñoz, F; Huerga, C; Aza, Z; Plaza-Núñez, R; Serrada, A; Bret-Zurita, M; Parrón, M; Prieto-Areyano, C; Garzón-Moll, G; Madero, R; Guibelalde, E.
Afiliación
  • Alejo L; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Corredoira E; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Sánchez-Muñoz F; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Huerga C; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Aza Z; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Plaza-Núñez R; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Serrada A; 1 Department of Medical Physics, La Paz University Hospital , Madrid , Spain.
  • Bret-Zurita M; 2 Department of Paediatric Radiology, La Paz University Hospital , Madrid , Spain.
  • Parrón M; 2 Department of Paediatric Radiology, La Paz University Hospital , Madrid , Spain.
  • Prieto-Areyano C; 2 Department of Paediatric Radiology, La Paz University Hospital , Madrid , Spain.
  • Garzón-Moll G; 2 Department of Paediatric Radiology, La Paz University Hospital , Madrid , Spain.
  • Madero R; 3 Department of Biostatistics, La Paz University Hospital , Madrid , Spain.
  • Guibelalde E; 4 Department of Radiology, Complutense University , Madrid , Spain.
Br J Radiol ; 91(1086): 20180022, 2018 Jun.
Article en En | MEDLINE | ID: mdl-29537302
Objective: The new 2013/59 EURATOM Directive (ED) demands dosimetric optimisation procedures without undue delay. The aim of this study was to optimise paediatric conventional radiology examinations applying the ED without compromising the clinical diagnosis. METHODS: Automatic dose management software (ADMS) was used to analyse 2678 studies of children from birth to 5 years of age, obtaining local diagnostic reference levels (DRLs) in terms of entrance surface air kerma. Given local DRL for infants and chest examinations exceeded the European Commission (EC) DRL, an optimisation was performed decreasing the kVp and applying the automatic control exposure. To assess the image quality, an analysis of high-contrast resolution (HCSR), signal-to-noise ratio (SNR) and figure of merit (FOM) was performed, as well as a blind test based on the generalised estimating equations method. RESULTS: For newborns and chest examinations, the local DRL exceeded the EC DRL by 113%. After the optimisation, a reduction of 54% was obtained. No significant differences were found in the image quality blind test. A decrease in SNR (-37%) and HCSR (-68%), and an increase in FOM (42%), was observed. CONCLUSION: ADMS allows the fast calculation of local DRLs and the performance of optimisation procedures in babies without delay. However, physical and clinical analyses of image quality remain to be needed to ensure the diagnostic integrity after the optimisation process. Advances in knowledge: ADMS are useful to detect radiation protection problems and to perform optimisation procedures in paediatric conventional imaging without undue delay, as ED requires.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Radiometría / Programas Informáticos / Radiografía Torácica Tipo de estudio: Guideline Límite: Child, preschool / Humans / Infant / Newborn Idioma: En Revista: Br J Radiol Año: 2018 Tipo del documento: Article País de afiliación: España Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Radiometría / Programas Informáticos / Radiografía Torácica Tipo de estudio: Guideline Límite: Child, preschool / Humans / Infant / Newborn Idioma: En Revista: Br J Radiol Año: 2018 Tipo del documento: Article País de afiliación: España Pais de publicación: Reino Unido