Reliability of automated synovial fluid cell counting with Mindray BC-6800 body fluid mode.
Buoro, S; Seghezzi, M; Manenti, B; Mecca, T; Candiago, E; Vidali, M; Ottomano, C; Castelli, C C; Ferrari, R; Zappalà, G; Lippi, G.
Afiliación
  • Buoro S; Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Seghezzi M; Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Manenti B; Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Mecca T; Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Candiago E; Pathology, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Vidali M; Immunohaematology and Transfusion Medicine Unit, SS Trinità Hospital, Borgomanero, Italy.
  • Ottomano C; Synlab, Monza, Italy.
  • Castelli CC; Orthopedics and Traumatology Unit, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Ferrari R; Orthopedics and Traumatology Unit, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Zappalà G; Orthopedics and Traumatology Unit, Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Lippi G; Section of Clinical Biochemistry, University of Verona, Verona, Italy.
Int J Lab Hematol ; 39(3): 337-346, 2017 Jun.
Article en En | MEDLINE | ID: mdl-28263031
INTRODUCTION: The enumeration and differentiation of nuclear elements in synovial fluid is a cornerstone for diagnosis and follow-up of many orthopedic and rheumatologic diseases. In this study, we evaluated the analytical performance of Mindray BC-6800 BF mode (BC-6800-BF) for synovial fluid analysis. METHODS: Overall, 78 synovial fluids were collected and analyzed with both BC-6800-BF and light microscopy. The study also entailed the assessment of limit of blank (LoB), limit of detection (LoD), limit of quantification (LoQ), carryover and linearity. RESULTS: The LoB for the parameters total cells and white blood cells was 6 × 106 cells/L, and the LoD and LoQ were instead 15 and 16 × 106 cells/L, respectively. Linearity was excellent and carryover was negligible. The agreement between BC-6800-BF and light microscopy was satisfactory for all samples pretreated with hyaluronidase, displaying a bias between -5.9% and 8.2%. CONCLUSIONS: The use of BC-6800-BF for synovial fluid analysis enables rapid and accurate assessment, especially for total cell and polymorphonuclear counts. The use of BC-6800-BF may therefore allow the replacement of optical analysis, especially in samples pretreated with hyaluronidase, thus allowing its routine use for the screening of synovial specimens.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Líquido Sinovial / Enfermedades Reumáticas / Citometría de Flujo Tipo de estudio: Evaluation_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Int J Lab Hematol Asunto de la revista: HEMATOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Líquido Sinovial / Enfermedades Reumáticas / Citometría de Flujo Tipo de estudio: Evaluation_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Int J Lab Hematol Asunto de la revista: HEMATOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Reino Unido