100 kHz Mueller polarimeter in reflection configuration.
Opt Lett
; 40(4): 645-8, 2015 Feb 15.
Article
en En
| MEDLINE
| ID: mdl-25680171
A new setup is proposed to perform high-speed Mueller polarimetry by spectral coding of polarization in a reflection configuration. The system uses a swept laser source and a photodiode, which results in a simple optical setup that allows measurement of Mueller matrices at 100 kHz repetition rate. A special focus is made on the influence of the cube beam splitter polarimetric response, which is essential to measurements in a reflection configuration. The instrument is first validated on reference samples for single-point measurements, and the effect of a proper system calibration is also demonstrated on polarimetric images. The device is intended to be implemented within a laser scanning microscope to perform multimodal imaging (confocal/multiphoton and Mueller polarimetry).
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Dispositivos Ópticos
Idioma:
En
Revista:
Opt Lett
Año:
2015
Tipo del documento:
Article
Pais de publicación:
Estados Unidos