Development of system using beam's eye view images to measure respiratory motion tracking errors in image-guided robotic radiosurgery system.
J Appl Clin Med Phys ; 16(1): 5049, 2015 Jan 08.
Article en En | MEDLINE | ID: mdl-25679160
The accuracy of the CyberKnife Synchrony Respiratory Tracking System (SRTS) is considered to be patient-dependent because the SRTS relies on an individual correlation between the internal tumor position (ITP) and the external marker position (EMP), as well as a prediction method to compensate for the delay incurred to adjust the position of the linear accelerator (linac). We aimed to develop a system for obtaining pretreatment statistical measurements of the SRTS tracking error by using beam's eye view (BEV) images, to enable the prediction of the patient-specific accuracy. The respiratory motion data for the ITP and the EMP were derived from cine MR images obtained from 23 patients. The dynamic motion phantom was used to reproduce both the ITP and EMP motions. The CyberKnife was subsequently operated with the SRTS, with a CCD camera mounted on the head of the linac. BEV images from the CCD camera were recorded during the tracking of a ball target by the linac. The tracking error was measured at 15 Hz using in-house software. To assess the precision of the position detection using an MR image, the positions of test tubes (determined from MR images) were compared with their actual positions. To assess the precision of the position detection of the ball, ball positions measured from BEV images were compared with values measured using a Vernier caliper. The SRTS accuracy was evaluated by determining the tracking error that could be identified with a probability of more than 95% (Ep95). The detection precision of the tumor position (determined from cine MR images) was < 0.2 mm. The detection precision of the tracking error when using the BEV images was < 0.2mm. These two detection precisions were derived from our measurement system and were not obtained from the SRTS. The median of Ep95 was found to be 1.5 (range, 1.0-3.5) mm. The difference between the minimum and maximum Ep95 was 2.5mm, indicating that this provides a better means of evaluating patient-specific SRTS accuracy. A suitable margin, based on the predicted patient-specific SRTS accuracy, can be added to the clinical target volume.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Robótica / Radiocirugia / Radioterapia de Intensidad Modulada / Técnicas de Imagen Sincronizada Respiratorias / Errores de Configuración en Radioterapia / Radioterapia Guiada por Imagen / Neoplasias Pulmonares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2015 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Robótica / Radiocirugia / Radioterapia de Intensidad Modulada / Técnicas de Imagen Sincronizada Respiratorias / Errores de Configuración en Radioterapia / Radioterapia Guiada por Imagen / Neoplasias Pulmonares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2015 Tipo del documento: Article Pais de publicación: Estados Unidos