OBJECTIVE: To develop an autologous total ossicular replacement prosthesis with sustainable hearing results. METHODS: The ears of 40 patients, who had chronic otitis media with absent suprastructure of the stapes and long process of the incus, were repaired using the autologous total ossicular replacement technique. Post-operative results were evaluated after 6 and 12 months on the basis of average pure tone air conduction and average air-bone gap measured at 0.5, 1, 2 and 3 kHz. RESULTS: Successful rehabilitation of pure tone average to 30 dB or less was achieved in 75 per cent of patients, and air-bone gap to 20 dB or less was attained in 82.5 per cent of patients. Overall mean improvement in air-bone gap was 23.9 ± 8.5 dB (p < 0.001). Mean improvements in air-bone gap were significantly greater (p < 0.05) in the tympanoplasty only group (27.3 ± 6.6 dB) and the intact canal wall tympanoplasty group (25.9 ± 6.3 dB) than in the canal wall down tympanoplasty group (16.3 ± 8.9 dB). CONCLUSION: This paper describes an autologous total ossicular replacement prosthesis that is biocompatible, stable, magnetic resonance imaging compatible and, above all, results in sustainable hearing improvement.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Otitis Media
/
Prótesis Osicular
/
Reemplazo Osicular
Límite:
Adolescent
/
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
J Laryngol Otol
Asunto de la revista:
OTORRINOLARINGOLOGIA
Año:
2014
Tipo del documento:
Article
País de afiliación:
India
Pais de publicación:
Reino Unido