Effect of frequency on entrance and propagation pattern of high-frequency (radiofrequency) electrical currents in biologic tissues.
Taheri, A; Mansoori, P; Huang, K E; Feldman, S R.
Afiliación
  • Taheri A; Center for Dermatology Research, Department of Dermatology, Wake Forest School of Medicine, Winston-Salem, NC, USA.
  • Mansoori P; Department of Pathology, Wake Forest School of Medicine, Winston-Salem, NC, USA.
  • Huang KE; Center for Dermatology Research, Department of Dermatology, Wake Forest School of Medicine, Winston-Salem, NC, USA.
  • Feldman SR; Center for Dermatology Research, Department of Dermatology, Wake Forest School of Medicine, Winston-Salem, NC, USA.
Skin Res Technol ; 21(3): 291-4, 2015 Aug.
Article en En | MEDLINE | ID: mdl-25329232
BACKGROUND: Radiofrequency electrical currents have a tendency to move toward the center of the bulk of biologic tissues. OBJECTIVES: To evaluate the effect of the frequency of currents on their entrance and propagation pattern in biologic tissues. MATERIALS AND METHODS: Three electrosurgical generators with 0.4, 1.5, and 3 MHz frequency outputs were studied. Current was applied using a metallic needle introduced into a piece of cow liver, with different amounts of energy delivered at multiple points. Cross-sections of the liver were then studied for tissue effect. The diameters of the coagulated areas at the deepest and most superficial parts were measured. The tendency of the currents for penetration in the deeper layers of tissue rather than in the superficial layers was assessed using the superficial diameter/deep diameter ratio. RESULTS: Diameter of coagulated area was larger around deeper parts than around superficial parts of the electrode. No correlation between frequency of current and the superficial/deep diameter ratio of the coagulation zone was found. CONCLUSION: Radiofrequency currents have a tendency to move toward the center of the tissue. Frequency of current over the range of 0.4-3 MHz did not show any effect on this tendency.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ablación por Catéter / Hepatectomía / Hígado Límite: Animals Idioma: En Revista: Skin Res Technol Asunto de la revista: DERMATOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ablación por Catéter / Hepatectomía / Hígado Límite: Animals Idioma: En Revista: Skin Res Technol Asunto de la revista: DERMATOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Reino Unido