Six-year in-vitro reliability results of the HeartWare HVAD pump.
Reyes, Carlos; Chorpenning, Katherine; LaRose, Jeffrey A; Gomez, Ramiro; Tamez, Daniel.
Afiliación
  • Reyes C; From the HeartWare, Inc., Pearland, Texas.
ASAIO J ; 60(5): 541-4, 2014.
Article en En | MEDLINE | ID: mdl-24814839
As a result of stagnant heart transplantation rates, ventricular assist devices (VADs) have become a widely accepted therapy for the treatment of advanced-stage heart failure. Long-term reliability of VADs will become increasingly vital as the population of destination therapy patients expands. In this study, eight HVAD pumps (n = 8) completed a 6-year reliability test in the HeartWare Life Cycle Testing System, an in-vitro mock circulatory loop that simulated physiologic pressures and flows. Cumulative runtime for the pumps was 2,408 ± 60 days. During this time, no device failures of any type occurred. These results strongly support the durability of the pump design.