Your browser doesn't support javascript.
loading
The effect of interlocking parallel screws in subcapital femoral-neck fracture fixation: a cadaver study.
Basso, Trude; Klaksvik, Jomar; Foss, Olav Andreas.
Afiliación
  • Basso T; Orthopaedic Research Centre, Trondheim University Hospital, Norway. Electronic address: trude.basso@ntnu.no.
  • Klaksvik J; Orthopaedic Research Centre, Trondheim University Hospital, Norway.
  • Foss OA; Orthopaedic Research Centre, Trondheim University Hospital, Norway; The Department of Neuroscience, Norwegian University of Science and Technology, Trondheim, Norway.
Clin Biomech (Bristol, Avon) ; 29(2): 213-7, 2014 Feb.
Article en En | MEDLINE | ID: mdl-24325974
BACKGROUND: A new locking-plate for femoral-neck fractures that provides angular stability to three screws in an inverted triangle configuration was evaluated. The plate is not fixed to the lateral cortex and therefore represents a new treatment principle. METHODS: Twelve pairs of cadaver femurs (mean T-score -1,95 (range -4,5-0)) with subcapital femoral-neck fractures angulating 60° were randomly allocated to fracture-fixation using either three individual screws or three interlocked screws. Subject-specific axial force and torque were applied by a hip simulator and three-dimensional migrations were recorded. The femurs underwent 10,000 cycles of simulated partial weight-bearing, followed by 10,000 cycles of simulated full weight-bearing and stair climbing. FINDINGS: On average interlocking reduced femoral-head centre migrations 1.6mm (95% CI 0.1-3.1, P = 0.04). The intra-pair correlation of migration was 0.953 (Pearson's r). Interlocking did not change rotational stability (P = 0.87). Adding a locking plate did not affect the risk of failure, however all failed femurs were fixed using the smallest-sized aiming guide. INTERPRETATIONS: Adding a lateral interlocking plate to three screws might improve the fracture stability. However, none of the implants were able to resist the unwanted deformation of the proximal femur. Regardless of the fixation, female sex, reduced bone quality and small sized femurs appear to increase risk of failure.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Placas Óseas / Tornillos Óseos / Fracturas del Cuello Femoral / Fijación Interna de Fracturas Tipo de estudio: Clinical_trials / Evaluation_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Biomech (Bristol, Avon) Asunto de la revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2014 Tipo del documento: Article Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Placas Óseas / Tornillos Óseos / Fracturas del Cuello Femoral / Fijación Interna de Fracturas Tipo de estudio: Clinical_trials / Evaluation_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Biomech (Bristol, Avon) Asunto de la revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2014 Tipo del documento: Article Pais de publicación: Reino Unido