Your browser doesn't support javascript.
loading
Effects of surface texturing on the performance of biocompatible UHMWPE as a bearing material during in vitro lubricated sliding/rolling motion.
López-Cervantes, Adrián; Domínguez-López, Iván; Barceinas-Sánchez, José Dolores Oscar; García-García, Adrián Luis.
Afiliación
  • López-Cervantes A; CICATA-IPN, Unidad Querétaro, Cerro Blanco No. 141, Col. Colinas del Cimatario, Querétaro, Qro.76090, México.
J Mech Behav Biomed Mater ; 20: 45-53, 2013 Apr.
Article en En | MEDLINE | ID: mdl-23455163
The effect of surface texturing on the performance of biocompatible ultra-high molecular weight polyethylene (UHMWPE) as a bearing material has been investigated using the kinematic range of motions reported for a knee-joint replacement. An experimental apparatus consisting of a ball and a disk rotating independently from each other was used to compare the performance of UHMWPE textured versus plain surfaces, under different combinations of sliding and rolling motion, better known as sliding-to-rolling ratio (SRR). Performance was evaluated through the coefficient of traction of a tribosystem comprising a steel ball on a flat UHMWPE disk and distilled water at 36°C, acting as lubricant. A square array of cavities with diameter D=0.397mm and center-to-center spacing of 1.5D was machined on UHMWPE disks. The experimental design considered two levels for cavity depth, D and D/2, and two for the applied load, 17 and 25N. The SRR was varied from 1 to 11% and the mean speed range was set from 5 to 55mm/s, covering the kinematics and contact pressure conditions of a sauntering cycle on a knee-joint replacement. Stribeck curves of the plain and textured surfaces were obtained and compared against one another. The results demonstrate that the proposed surface pattern reduces the coefficient of traction of the tribological system for the 17N load in the entire kinematic range explored, while for the 25N load the effects were more noticeable at low mean speed and SRR, corresponding to the beginning of motion.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polietilenos / Materiales Biocompatibles / Prótesis de la Rodilla Límite: Humans Idioma: En Revista: J Mech Behav Biomed Mater Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2013 Tipo del documento: Article Pais de publicación: Países Bajos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polietilenos / Materiales Biocompatibles / Prótesis de la Rodilla Límite: Humans Idioma: En Revista: J Mech Behav Biomed Mater Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2013 Tipo del documento: Article Pais de publicación: Países Bajos