Your browser doesn't support javascript.
loading
Visual simulation through an aspheric aberration-correcting intraocular lens in subjects with different corneal profiles using adaptive optics.
Ruiz-Alcocer, Javier; Madrid-Costa, David; García-Lázaro, Santiago; Albarrán-Diego, César; Ferrer-Blasco, Teresa.
Afiliación
  • Ruiz-Alcocer J; Optics and Optometry Department, Universidad Europea de Madrid, Madrid, Spain. j.ruizalcocer@hotmail.com
Clin Exp Optom ; 96(4): 379-84, 2013 Jul.
Article en En | MEDLINE | ID: mdl-23298381
BACKGROUND: The aim of this study was to analyse the visual quality of the AcrySof IQ SN60WF(®) intraocular lens (IOL) when combined with different corneal profiles. METHODS: Ten eyes of 10 participants with no prior history of refractive or cataract surgery were evaluated. An adaptive optics visual simulator was used to simulate the wavefront aberration pattern of an aspheric aberration-correcting IOL (AcrySof IQ SN60WF(®)). Normal corneas (group A), low and high myopic corneal ablations (groups B and C, respectively) and low and high hyperopic corneal ablations (groups D and E, respectively) were also simulated. Monocular distance visual acuities at 100, 50 and 10 per cent of contrast were measured. RESULTS: At 100, 50 and 10 per cent contrast, no differences were found between groups A and B (p > 0.06 for all contrasts). Group A obtained better values than groups C, D and E for all contrasts (p = 0.031, p = 0.038, p = 0.032 at 100, 50 and 10 per cent of contrast, respectively). At the same time, group B obtained better values than groups C, D and E (p = 0.041, p = 0.042, p = 0.036 at 100, 50 and 10 per cent of contrast, respectively). Within the five groups, the worst results were always obtained for group E (p = 0.017, p = 0.021 and p = 0.025 at 100, 50 and 10 per cent of contrast, respectively). CONCLUSIONS: The results suggest that the aspheric aberration-correcting IOL studied provides comparable results, when it is combined with normal corneas and with corneas with simulated low myopic ablations. When negative amounts of residual spherical aberration after cataract surgery are expected to be achieved, IOLs with more positive spherical aberration should be considered.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantación de Lentes Intraoculares / Aberración de Frente de Onda Corneal Límite: Adult / Humans Idioma: En Revista: Clin Exp Optom Asunto de la revista: OPTOMETRIA Año: 2013 Tipo del documento: Article País de afiliación: España Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantación de Lentes Intraoculares / Aberración de Frente de Onda Corneal Límite: Adult / Humans Idioma: En Revista: Clin Exp Optom Asunto de la revista: OPTOMETRIA Año: 2013 Tipo del documento: Article País de afiliación: España Pais de publicación: Estados Unidos