Your browser doesn't support javascript.
loading
A comparative study of color quenching correction methods for Cerenkov counting.
Tsroya, S; Pelled, O; German, U; Katorza, E; Abraham, A; Alfassi, Z B.
Afiliación
  • Tsroya S; Nuclear Research Center Negev, Beer Sheva, Israel.
Appl Radiat Isot ; 70(2): 397-403, 2012 Feb.
Article en En | MEDLINE | ID: mdl-22041298
The color quenching correction for Cerenkov counting is normally based on indexes defined for liquid scintillation spectra. Because of the differences in the spectrum shapes, these indexes are not optimal, and a different method was proposed based on the integral counting of the external (152)Eu source spectra of a 1200 Quantulus system. We have shown that this method, named ESAR (external source area ratio), is superior to all the other correction methods, offering the highest sensitivity and the widest range of application.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2012 Tipo del documento: Article País de afiliación: Israel Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2012 Tipo del documento: Article País de afiliación: Israel Pais de publicación: Reino Unido