First permanent implant of nerve stimulation leads activated by surface electrodes, enabling hand grasp and release: the stimulus router neuroprosthesis.
Gan, Liu Shi; Ravid, Einat; Kowalczewski, Jan Andrzej; Olson, Jaret Lawrence; Morhart, Michael; Prochazka, Arthur.
Afiliación
  • Gan LS; University of Alberta, Edmonton, Alberta, Canada.
Neurorehabil Neural Repair ; 26(4): 335-43, 2012 May.
Article en En | MEDLINE | ID: mdl-21959674
BACKGROUND: . More than 150 000 neuroprostheses (NPs) have been implanted in people to restore bodily function in a variety of neural disorders. The authors developed a novel NP, the Stimulus Router System (SRS), in which only passive leads are implanted. Each lead picks up a portion of the current delivered through the skin by an external stimulator. OBJECTIVE: . The authors report on the first human implant of an SRS. METHODS: . The recipient was a tetraplegic man with bilateral hand paralysis. Three SRS leads were implanted in his right forearm to activate the finger extensors, finger flexors, and thumb flexor. A wristlet containing a surface stimulator and electrodes was used to pass trains of electrical pulses through the skin to each lead. Hand opening and grasp were controlled via a wireless earpiece that sensed small tooth-clicks and transmitted signals to the wristlet. RESULTS: . The current required to activate the muscles was less than half that required prior to implantation and below perceptual threshold. Maximal grip force and hand opening aperture were both larger using the SRS. The implanted leads have remained functional for 3 years. The recipient reported various tasks of daily life that improved during SRS usage. An electronic counter revealed mean monthly usage of 18.5 hours, equivalent to 55 hours of continuous manual activity. CONCLUSIONS: . This first implant of the SRS indicates that it can be effective and reliable and has potential to provide an alternative to existing NPs.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cuadriplejía / Traumatismos de la Médula Espinal / Terapia por Estimulación Eléctrica / Fuerza de la Mano / Electrodos Implantados Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies Límite: Humans / Male / Middle aged Idioma: En Revista: Neurorehabil Neural Repair Asunto de la revista: NEUROLOGIA / REABILITACAO Año: 2012 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cuadriplejía / Traumatismos de la Médula Espinal / Terapia por Estimulación Eléctrica / Fuerza de la Mano / Electrodos Implantados Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies Límite: Humans / Male / Middle aged Idioma: En Revista: Neurorehabil Neural Repair Asunto de la revista: NEUROLOGIA / REABILITACAO Año: 2012 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos