Evaluation of portable CT scanners for otologic image-guided surgery.
Int J Comput Assist Radiol Surg
; 7(2): 315-21, 2012 Mar.
Article
en En
| MEDLINE
| ID: mdl-21779768
PURPOSE: Portable CT scanners are beneficial for diagnosis in the intensive care unit, emergency room, and operating room. Portable fixed-base versus translating-base CT systems were evaluated for otologic image-guided surgical (IGS) applications based on geometric accuracy and utility for percutaneous cochlear implantation. METHODS: Five cadaveric skulls were fitted with fiducial markers and scanned using both a translating-base, 8-slice CT scanner (CereTom(®)) and a fixed-base, flat-panel, volume CT (fpVCT) scanner (Xoran xCAT(®)). Images were analyzed for: (a) subjective quality (i.e., noise), (b) consistency of attenuation measurements (Hounsfield units) across similar tissue, and (c) geometric accuracy of fiducial marker positions. The utility of these scanners in clinical IGS cases was tested. RESULTS: Five cadaveric specimens were scanned using each of the scanners. The translating-base, 8-slice CT scanner had spatially consistent Hounsfield units, and the image quality was subjectively good. However, because of movement variations during scanning, the geometric accuracy of fiducial marker positions was low. The fixed-base, fpVCT system had high spatial resolution, but the images were noisy and had spatially inconsistent attenuation measurements, while the geometric representation of the fiducial markers was highly accurate. CONCLUSION: Two types of portable CT scanners were evaluated for otologic IGS. The translating-base, 8-slice CT scanner provided better image quality than a fixed-base, fpVCT scanner. However, the inherent error in three-dimensional spatial relationships by the translating-based system makes it suboptimal for otologic IGS use.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Procedimientos Quirúrgicos Otológicos
/
Sistemas de Atención de Punto
/
Tomografía Computarizada Multidetector
/
Oído Interno
Límite:
Humans
Idioma:
En
Revista:
Int J Comput Assist Radiol Surg
Asunto de la revista:
RADIOLOGIA
Año:
2012
Tipo del documento:
Article
País de afiliación:
Estados Unidos
Pais de publicación:
Alemania