Measurement of radon concentration in water using the portable radon survey meter.
Yokoyama, S; Mori, N; Shimo, M; Fukushi, M; Ohnuma, S.
Afiliación
  • Yokoyama S; Faculty of Medical Technology, School of Health Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi 470-1192, Japan. sumi0704@fujita-hu.ac.jp
Radiat Prot Dosimetry ; 146(1-3): 23-6, 2011 Jul.
Article en En | MEDLINE | ID: mdl-21493609
A measurement method for measuring radon in water using the portable radon survey meter (RnSM) was developed. The container with propeller was used to stir the water samples and release radon from the water into the air in a sample box of the RnSM. In this method, the measurement of error would be <20 %, when the radon concentration in the mineral water was >20 Bq l(-1).
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Contaminantes Radiactivos del Agua / Monitoreo de Radiación / Radón Idioma: En Revista: Radiat Prot Dosimetry Año: 2011 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Reino Unido