Influence of rotary blood pumps over preload recruitable stroke work.
Annu Int Conf IEEE Eng Med Biol Soc
; 2010: 2367-70, 2010.
Article
en En
| MEDLINE
| ID: mdl-21097228
When recovery of the cardiac function is detected in assisted hearts, the ventricular assist device can be removed. Due to the invasiveness of the surgical procedure, an accurate assessment of cardiac function is fundamental for the treatment success. The main challenge for the detection of cardiac function during assistance is to know whether the cardiac function index represents the cardiac function after pump removal independently of the pump assist rate. Therefore in this paper we present an evaluation of the influence of the pump over the slope of the preload recruitable stroke work, a cardiac function index. Analyzing data from four acute animal experiments, we found that the pump affects the stroke work, which could be corrected by the end diastolic volume. However, the data set examined was limited and further investigation is necessary.