Development of a time-resolved soft x-ray spectrometer for laser produced plasma experiments.
Cone, K V; Dunn, J; Schneider, M B; Baldis, H A; Brown, G V; Emig, J; James, D L; May, M J; Park, J; Shepherd, R; Widmann, K.
Afiliación
  • Cone KV; Lawrence Livermore National Laboratory, Livermore, California 94551, USA. cone2@llnl.gov
Rev Sci Instrum ; 81(10): 10E318, 2010 Oct.
Article en En | MEDLINE | ID: mdl-21034016
A 2400 lines/mm variable-spaced grating spectrometer has been used to measure soft x-ray emission (8-22 Å) from laser-produced plasma experiments at Lawrence Livermore National Laboratory's Compact Multipulse Terrawatt (COMET) Laser Facility. The spectrometer was coupled to a Kentech x-ray streak camera to study the temporal evolution of soft x rays emitted from the back of the Mylar and the copper foils irradiated at 10(15) W/cm(2). The instrument demonstrated a resolving power of ∼120 at 19 Å with a time resolution of 31 ps. The time-resolved copper emission spectrum was consistent with a photodiode monitoring the laser temporal pulse shape and indicated that the soft x-ray emission follows the laser heating of the target. The time and spectral resolutions of this diagnostic make it useful for studies of high temperature plasmas.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos