Your browser doesn't support javascript.
loading
Accurate determination of arsenic in arsenobetaine standard solutions of BCR-626 and NMIJ CRM 7901-a by neutron activation analysis coupled with internal standard method.
Miura, Tsutomu; Chiba, Koichi; Kuroiwa, Takayoshi; Narukawa, Tomohiro; Hioki, Akiharu; Matsue, Hideaki.
Afiliación
  • Miura T; National Metrology Institute of Japan, AIST, Tsukuba Central 3, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan.
Talanta ; 82(4): 1143-8, 2010 Sep 15.
Article en En | MEDLINE | ID: mdl-20801310
Neutron activation analysis (NAA) coupled with an internal standard method was applied for the determination of As in the certified reference material (CRM) of arsenobetaine (AB) standard solutions to verify their certified values. Gold was used as an internal standard to compensate for the difference of the neutron exposure in an irradiation capsule and to improve the sample-to-sample repeatability. Application of the internal standard method significantly improved linearity of the calibration curve up to 1 microg of As, too. The analytical reliability of the proposed method was evaluated by k(0)-standardization NAA. The analytical results of As in AB standard solutions of BCR-626 and NMIJ CRM 7901-a were (499+/-55)mgkg(-1) (k=2) and (10.16+/-0.15)mgkg(-1) (k=2), respectively. These values were found to be 15-20% higher than the certified values. The between-bottle variation of BCR-626 was much larger than the expanded uncertainty of the certified value, although that of NMIJ CRM 7901-a was almost negligible.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Talanta Año: 2010 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Países Bajos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Talanta Año: 2010 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Países Bajos