Improved performance characteristics of the von Willebrand factor ristocetin cofactor activity assay using a novel automated assay protocol.
Hillarp, A; Stadler, M; Haderer, C; Weinberger, J; Kessler, C M; Römisch, J.
Afiliación
  • Hillarp A; Malmö Centre for Thrombosis and Haemostasis, University and Regional Laboratories Region Scania, Malmö University Hospital, Malmö, Sweden. andreas.hillarp@med.lu.se
J Thromb Haemost ; 8(10): 2216-23, 2010 Oct.
Article en En | MEDLINE | ID: mdl-20727070
UNLABELLED: BACKGROUND, OBJECTIVES AND METHODS: An accurate, sensitive and precise assay for reliable determination of the ristocetin cofactor activity of von Willebrand factor (VWF:RCo) in plasma and von Willebrand Factor (VWF)-containing concentrates has been evaluated. The assay is based on a commercially available automated protocol with modifications including a combination of adding additional ristocetin and the use of two calibration curves for the high and low measuring ranges. RESULTS: Addition of extra ristocetin resulted in improved measurement of VWF recoveries from various VWF-containing concentrates that were underestimated using the standard automated protocol. The modifications resulted in improved assay performance over an extended measuring range (2.00-0.03 IUmL(-1) ). Accuracy was tested using VWF deficiency plasma spiked with the 1st international standard (IS) for VWF concentrate. Seven dilutions, ranging from 1.80 to 0.05IUmL(-1) , were analyzed and resulted in measured concentrations between 80% and 100% of the assigned potency of the standard. Linearity was determined from the regression plot of the same concentrate dilutions and resulted in a correlation coefficient of 0.998. The repeatability, expressed as coefficient of variation, was 2% in the normal range (0.90IUmL(-1) ) and 8% at the level of 0.05IUmL(-1) . The corresponding reproducibility results were 2% and 15% at the normal and low measuring ranges, respectively. CONCLUSIONS: Analysis of patients with von Willebrand disease (VWD) indicates that the modified automated BCS(®) protocol has a superior discrimination power compared with the standard protocol. This is especially true in samples with low VWF, as in patients with type 3 VWD.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedades de von Willebrand / Pruebas de Coagulación Sanguínea / Ristocetina Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Thromb Haemost Asunto de la revista: HEMATOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedades de von Willebrand / Pruebas de Coagulación Sanguínea / Ristocetina Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Thromb Haemost Asunto de la revista: HEMATOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Reino Unido