Evaluation of radiation doses from MDCT-imaging in otolaryngology.
Radiat Prot Dosimetry
; 136(1): 38-44, 2009 Aug.
Article
en En
| MEDLINE
| ID: mdl-19617599
The purpose of this study was to clarify patient doses in the current otolaryngological multi-detector row computed tomography (MDCT) examinations. Patient doses were measured with an in-phantom dosimetry system which was composed of 48 photodiode dosimeters embedded within an anthropomorphic phantom. Organ and effective doses were evaluated according to the International Commission on Radiological Protection Publication 103. In neck CT, doses for salivary glands and for thyroid were high, 7.6-29.9 and 13.4-60.3 mGy, respectively. In sinus CT, brain and lens doses were high, 7.6-24.6 and 10.6-32.0 mGy, respectively, and in inner ear CT, lens dose was 8.0-35.3 mGy. Effective doses were 1.8-6.6 mSv in neck CT, 0.5-0.9 mSv in sinus CT and 0.3-0.6 mSv in inner ear CT. The present dose data would be used to estimate radiation risks for patients undergoing otolaryngological MDCT examinations.