Color quench correction for low level Cherenkov counting.
Appl Radiat Isot
; 67(5): 805-8, 2009 May.
Article
en En
| MEDLINE
| ID: mdl-19249219
The Cherenkov counting efficiency varies strongly with color quenching, thus correction curves must be used to obtain correct results. The external (152)Eu source of a Quantulus 1220 liquid scintillation counting (LSC) system was used to obtain a quench indicative parameter based on spectra area ratio. A color quench correction curve for aqueous samples containing (90)Sr/(90)Y was prepared. The main advantage of this method over the common spectra indicators is its usefulness also for low level Cherenkov counting.