A real-time ultrasound time-domain correlation blood flowmeter. II. Performance and experimental verification.
Hein, I A; O'Brien, W R.
Afiliación
  • Hein IA; Dept. of Electr. and Comput. Eng., Illinois Univ., Urbana, IL.
Article en En | MEDLINE | ID: mdl-18263246
For Part I, see ibid., vol.40, no.6, p.768-775 (1993). The performance of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator under various flow conditions for different correlation parameter settings is evaluated. Previous work has incorporated a weighted-averaging scheme to estimate the flow velocity over a wide number of echo signal spacings. The present evaluation indicates that a linear averaging of 50 adjacent echo signal pairs produces equivalent accuracy and precision at real-time processing speed as compared to previous non-real time systems incorporating weighted averaging.
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: IEEE Trans Ultrason Ferroelectr Freq Control Asunto de la revista: MEDICINA NUCLEAR Año: 1993 Tipo del documento: Article Pais de publicación: Estados Unidos