A real-time ultrasound time-domain correlation blood flowmeter. II. Performance and experimental verification.
IEEE Trans Ultrason Ferroelectr Freq Control
; 40(6): 776-85, 1993.
Article
en En
| MEDLINE
| ID: mdl-18263246
For Part I, see ibid., vol.40, no.6, p.768-775 (1993). The performance of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator under various flow conditions for different correlation parameter settings is evaluated. Previous work has incorporated a weighted-averaging scheme to estimate the flow velocity over a wide number of echo signal spacings. The present evaluation indicates that a linear averaging of 50 adjacent echo signal pairs produces equivalent accuracy and precision at real-time processing speed as compared to previous non-real time systems incorporating weighted averaging.