The electrical properties of the solid state/fluid (Ringer solution) interface for phosphorous- and boron-doped porous silicon are reported and the benefits of using porous silicon as neural recording electrodes are discussed. The impedance, reactance and resistance for doped porous and planar silicon, in Ringer solution, were compared to gold electrodes. Planar silicon displayed approximately a three times higher reactance than porous electrodes. The phosphorous-doped porous electrodes displayed a similar reactance compared to the gold electrodes.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fósforo / Silicio / Ingeniería Biomédica / Boro Límite: Animals / Humans Idioma: En Revista: J Biomater Sci Polym Ed Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2007 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Reino Unido
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fósforo / Silicio / Ingeniería Biomédica / Boro Límite: Animals / Humans Idioma: En Revista: J Biomater Sci Polym Ed Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2007 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Reino Unido