Intercomparison of ionisation chamber measurements from (125)I seeds.
Appl Radiat Isot
; 65(5): 517-8, 2007 May.
Article
en En
| MEDLINE
| ID: mdl-17258464
The reference air kerma rates of a set of individual (125)I seeds were calculated from current measurements of a calibrated re-entrant ionisation chamber. Single seeds were distributed to seven Australian brachytherapy centres for the same measurement with the user's instrumentation. Results are expressed as the ratio of the reference air kerma rate measured by the Australian Nuclear Science & Technology Organisation (ANSTO) to the reference air kerma rate measured at the centre. The intercomparison ratios of all participants were within +/-5% of unity.