Determination of quality parameters from statistical analysis of routine TLD dosimetry data.
Radiat Prot Dosimetry
; 119(1-4): 306-9, 2006.
Article
en En
| MEDLINE
| ID: mdl-16709715
Following the as low as reasonably achievable (ALARA) practice, there is a need to measure very low doses, of the same order of magnitude as the natural background, and the limits of detection of the dosimetry systems. The different contributions of the background signals to the total zero dose reading of thermoluminescence dosemeter (TLD) cards were analysed by using the common basic definitions of statistical indicators: the critical level (L(C)), the detection limit (L(D)) and the determination limit (L(Q)). These key statistical parameters for the system operated at NRC-Negev were quantified, based on the history of readings of the calibration cards in use. The electronic noise seems to play a minor role, but the reading of the Teflon coating (without the presence of a TLD crystal) gave a significant contribution.