The low energy X-ray response of the LiF:Mg:Cu:P thermoluminescent dosemeter: a comparison with LiF:Mg:Ti.
Br J Radiol
; 78(930): 543-7, 2005 Jun.
Article
en En
| MEDLINE
| ID: mdl-15900061
LiF:Mg:Cu:P thermoluminescent dosemeters (TLD) can be used for the same X-ray dosimetry applications as LiF:Mg:Ti, with each type having the disadvantage of a response dependent on energy, particularly at low energies. Measurements were made of the response per unit air kerma of LiF:Mg:Cu:P and LiF:Mg:Ti to nine quasi-monoenergetic X-ray beams with mean energies from 12 keV to 208 keV. Each measurement was normalized to the value produced by 6 MV X-rays. LiF:Mg:Cu:P was found to under-respond to a majority of these radiations whereas LiF:Mg:Ti over-responded to a majority. Their smallest relative measured response was produced by the lowest energy beam, and the maximum measured relative response of 1.15+/-0.07 and 1.21+/-0.07 for LiF:Mg:Cu:P and LiF:Mg:Ti, respectively, occurred at 33 keV. Energy response coefficients were derived from these measurements to estimate the error introduced by using either type of TLD to measure the dose from an X-ray spectrum different to that used for its absolute response calibration. It was calculated that if the response of either type of TLD was calibrated at 100 kVp, then an error of no more than +/-2% would be introduced into measurements of tube output at potentials of 50-130 kVp. LiF:Mg:Cu:P was found to introduce a larger error (up to 30%) into the measurement of body exit dose than LiF:Mg:Ti at tube potentials of 40-150 kVp, if its absolute response was calibrated using the corresponding body entrance beam. The method should allow this type of error to be estimated in other dosimetry applications for either type of TLD.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Dosimetría Termoluminiscente
/
Rayos X
Límite:
Humans
Idioma:
En
Revista:
Br J Radiol
Año:
2005
Tipo del documento:
Article
Pais de publicación:
Reino Unido