Speciation of arsenic in water samples by high-performance liquid chromatography-hydride generation-atomic absorption spectrometry at trace levels using a post-column reaction system.
Stummeyer, J; Harazim, B; Wippermann, T.
Afiliación
  • Stummeyer J; Bundesanstalt für Geowissenschaften und Rohstoffe, Stilleweg 2, D-30655, Hannover, Germany.
Anal Bioanal Chem ; 354(3): 344-51, 1996 Jan.
Article en En | MEDLINE | ID: mdl-15048460
Anion-exchange HPLC has been combined with hydride generation - atomic absorption spectrometry (HG-AAS) for the routine speciation of arsenite, arsenate, monomethylarsenic acid and dimethylarsinic acid. The sensitivity of the AAS-detection was increased by a post-column reaction system to achieve complete formation of volatile arsines from the methylated species and arsenate. The system allows the quantitative determination of 0.5 microg/l of each arsenic compound in water samples. The stability of synthetical and natural water containing arsenic at trace levels was investigated. To preserve stored water samples, a method for quantitative separation of arsenate at high pH-values with the basic anion-exchange resin Dowex 1x8 was developed.
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Anal Bioanal Chem Año: 1996 Tipo del documento: Article País de afiliación: Alemania Pais de publicación: Alemania