Response of a tissue equivalent proportional counter to neutrons.
Badhwar, G D; Robbins, D E; Gibbons, F; Braby, L A.
Afiliación
  • Badhwar GD; NASA Johnson Space Center, Houston, TX 77058-3696, USA.
Radiat Meas ; 35(6): 551-6, 2002 Dec.
Article en En | MEDLINE | ID: mdl-12455500
The absorbed dose as a function of lineal energy was measured at the CERN-EC Reference-field Facility (CERF) using a 512-channel tissue equivalent proportional counter (TEPC), and neutron dose equivalent response evaluated. Although there are some differences, the measured dose equivalent is in agreement with that measured by the 16-channel HANDI tissue equivalent counter. Comparison of TEPC measurements with those made by a silicon solid-state detector for low linear energy transfer particles produced by the same beam, is presented. The measurements show that about 4% of dose equivalent is delivered by particles heavier than protons generated in the conducting tissue equivalent plastic.
Asunto(s)
Palabras clave
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Monitoreo de Radiación / Nave Espacial / Transferencia Lineal de Energía / Neutrones Idioma: En Revista: Radiat Meas Asunto de la revista: RADIOLOGIA Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Reino Unido
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Monitoreo de Radiación / Nave Espacial / Transferencia Lineal de Energía / Neutrones Idioma: En Revista: Radiat Meas Asunto de la revista: RADIOLOGIA Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Reino Unido