[Evaluation of Emit netilmicin and amikacin assays on Dimension RXL HM (Dade Behring) open channels]. / Adaptation des trousses Emit nétilmicine et amikacine sur canaux ouverts du Dimension RXL HM (Dade Behring).
Ann Biol Clin (Paris)
; 60(6): 723-30, 2002.
Article
en Fr
| MEDLINE
| ID: mdl-12446241
The aim of this study was to evaluate and to validate an enzyme immunoassay in homogeneous phase for netilmicin and amikacin, adapted on the Dimension RXL HM (Dade Behring) machine. The results were compared with those obtained with automated polarization of fluorescence immunoassay using TDx FLx (Abbott). The protocol of the study and the analytical criteria were inspired by the protocol Valtec version 2002 recommended by the French Society of Clinical Biology (SFBC). The validation of this technique as adapted to the Dimension RXL HM has allowed its use for routine dosage adjustment of amikacin and netilmicin. The practicability is however the weak point of the adaptation of these techniques, even limiting as for their implementation.