Asymmetrical gradient coil for head imaging.
Tomasi, D; Xavier, R F; Foerster, B; Panepucci, H; Tannús, A; Vidoto, E L.
Afiliación
  • Tomasi D; Medical Department, Brookhaven National Laboratory, Upton, New York 11973, USA. tomasi@bnl.gov
Magn Reson Med ; 48(4): 707-14, 2002 Oct.
Article en En | MEDLINE | ID: mdl-12353289
This work presents a novel approach to develop dedicated transverse gradient coils for head imaging. The proposed coil design is based on the stochastic optimization of an asymmetrical stream function and improves the matching between the region-of-interest and the homogeneous gradient volume. Additionally, the electric field produced by these asymmetrical coils is 30% lower than that produced by standard symmetrical designs, which minimizes the risk of magnetostimulation of nerves in fast imaging techniques. A prototype of the asymmetrical gradient coil was built to test the method and magnetic field produced by the prototype was measured. Magnetic field measurements and electrical parameters of coils are in good agreement with theoretical calculations.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Cabeza Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Cabeza Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos