Comparación del tonómetro de Goldmann, neumotonómetro de contacto y el efecto del grosor corneal / Comparison of Goldmann tonometry, pneumotonometry and the effect of the central corneal thickness
Molina, N; Milla, E; Bitrian, E; Larena, C; Martínez, L.
Afiliación
  • Molina, N; Hospital Clinic i Provincial de Barcelona. Instituto Clinic de Oftalmología. Barcelona. España
  • Milla, E; Hospital Clinic i Provincial de Barcelona. Instituto Clinic de Oftalmología. Barcelona. España
  • Bitrian, E; Hospital Clinic i Provincial de Barcelona. Instituto Clinic de Oftalmología. Barcelona. España
  • Larena, C; Hospital Clinic i Provincial de Barcelona. Instituto Clinic de Oftalmología. Barcelona. España
  • Martínez, L; Hospital Clinic i Provincial de Barcelona. Instituto Clinicabout:blank de Oftalmología. Barcelona. España
Arch. Soc. Esp. Oftalmol ; 85(10): 325-328, oct. 2010. tab, graf
Article en Es | IBECS | ID: ibc-89418
Biblioteca responsable: ES1.1
Ubicación: BNCS
RESUMEN
Propósito: Comparar la medición de la presióintraocular (PIO) usando el ton¨®metro de aplanaci¨®n de Goldmann (GAT) y el neumotonómetro de contacto (NTC), y evaluar el efecto del grosor corneal central (CCT) en las mediciones. Método: Estudio prospectivo, clínico observacional. Se incluyeron 96 ojos de 49 pacientes 40 glaucomatosos, 29 hipertensos oculares y 27 controles. La PIO se midió con cada uno de los tonómetros. Además se midió el CCT. Resultados: Se observó un coeficiente de correlación fuerte entre GAT y el NTC (Pearson=0,922, p<0,001). La media de las diferencias pareadas entre GAT y el NTC fue de 0,72¡À2,82mmHg. La media de las diferencias entre GAT y el NTC fue menor a valores bajos de PIO y más alta a valores mayores de PIO. Las mediciones con GAT demostraron un aumento medio de la PIO de 0,47 mmHg por cada 10-Ìm de CCT, y el NTC 0,34 mmHg por cada 10¦Ìm, aunque la diferencia no fue estadísticamente significativa. Conclusiones: La neumotonometría de contacto es un método fiable en el diagnóstico y seguimiento de nuestra población. El efecto del CCT fue mayor para el tonómetro de Goldmann(AU)
ABSTRACT
PurposeTo compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT) and contact pneumotonometer (NTC), and to evaluate the effects of central corneal thickness (CCT) on the readings.MethodsThis was a prospective, observational clinical study, which included 96 eyes of 49 patients, 40 glaucomatous, 29 ocular hypertensive and 27 normal subjects. The IOP measurements were performed with each tonometer. The CCT was also measured.ResultsA strong correlation coefficient between GAT and the NTC was found (Pearson=0.922, p<0.001). The mean of paired differences between GAT and NTC was 0.72¡À2.82 mmHg. The mean difference between GAT and NTC was low at lower levels of PIO and higher at higher levels of IOP. Readings with the GAT showed a mean increase in IOP of 0.47 mmHg for every 10¦Ìm of CCT, and the NTC 0.34 mmHg for every 10-Ìm, the diference was not statistically significant.ConclusionsThe contact pneumotonometer is a reliable method in diagnosis and follow up of our population. The effect of the central corneal thickness was more important for the Goldmann applanation tonometer(AU)